The UWS Cloud Portal brings together system information, measurement reports, and system notifications in a single centralized platform for the first time.

  • Centralized platform for all system data, measurement reports, and system notifications
  • Automatic, guideline-compliant documentation of all WaterBoy and device data
  • Real-time notifications for deviations, leakages, and maintenance requirements
Log In Create New Account
EN
Cloud Portal
UWS Cloud Portal

The UWS Cloud Portal brings together system information, measurement reports, and system notifications in a single centralized platform for the first time.

  • Centralized platform for all system data, measurement reports, and system notifications
  • Automatic, guideline-compliant documentation of all WaterBoy and device data
  • Real-time notifications for deviations, leakages, and maintenance requirements
Log In Create New Account
EN
UWS – Your partner for cooling and heating water treatment
UWS – Your partner for cooling and heating water treatment
EN
  • Home
  • Calibration WaterBoy

Calibration WaterBoy

   

WaterBoy richtig kalibrieren

Our WaterBoy dual measuring device must be calibrated regularly in order to provide accurate values for the VDI 2035, ÖNORM H 5195-1 and SWKI BT 102-01 standards:

  • pH calibration: every 5 days (pH 7 and pH 10 mandatory according to VDI 2035, pH 4 optional)
  • Conductivity calibration: every 14 days (1413 µS/cm and 84 µS/cm)
  • UWS recommendation: Calibrate the measuring device at both conductivity points to achieve even more accurate measurement results.
  • However, for a valid calibration, it is sufficient to calibrate the conductivity at one of the two points.

The WaterBoy app helps you with precise instructions on how to proceed correctly during calibration.

Perform calibration with the UWS WaterBoy app

To do this, switch to the desired tab (μS or pH)
If the calibration is invalid or has expired, the message ‘Device must be calibrated’ is highlighted in red at the bottom.
If the calibration has expired, please recalibrate the respective measuring probe.
Click on ‘Calibrate’ for step-by-step instructions on calibration.
Important: As soon as the device has been calibrated to a calibration solution, the app jumps out of calibration again. To calibrate another point, click on ‘Calibrate’ again and follow the instructions.
The date of the calibration and the calibrated points are displayed after successful calibration.

A detailed description of how to calibrate the UWS WaterBoy without the app can be found here in the instructions.

The following calibration liquids are available:

Calibration fluid Conductivity
Article Article content Species number
Conductivity solution 84µS / cm 120 ml bottle 100030-8
Conductivity solution 1413 µS/cm 25 bags of 20 ml 100030-2
Conductivity solution 1413 µS/cm 500 ml bottle 100030-25
Calibration liquid pH
Buffer solution pH 4.01 25 bags of 20 ml 100030-3
Buffer solution pH 4.01 500 ml bottle 100030-30
Buffer solution PH 7 25 bags of 20 ml 100030-1
Buffer solution PH 7 500 ml bottle 100030-15
Buffer solution pH 10.01 25 bags of 20 ml 100030-7
Buffer solution pH 10.01 500 ml bottle 100030-70
Cleaning and storage solution
Storage solution for electrodes 25 ml bottle 100125
Storage solution for electrodes 230 ml bottle 100145
Storage solution for electrodes 500 ml bottle 100135
Cleaning solution for electrodes 25 bags of 20 ml 100030-6
Cleaning solution for electrodes 500 ml bottle 100136
Replacement set “WaterBoy” measuring case 10 pcs. 100153-1

We also offer the following measuring and test equipment:

article Item number
Measuring set for measuring water hardness 100032
Chloride test kit 100130
Test cutlery iron 100131
Nitrate test kit 100133
Sulphate test kit 100134
Measuring cup 50 ml 100030-4