Calibration WaterBoy

Only with the correct calibration and appropriate calibration liquids can you obtain accurate measured values for pH value and conductivity!

product-image

Calibrate correctly

Our WaterBoy dual measuring device must be calibrated regularly in order to provide accurate values for the VDI 2035, ÖNORM H 5195-1 and SWKI BT 102-01 standards:

  • pH calibration: every 5 days
  • Conductivity calibration: every 14 days

The WaterBoy app helps you with precise instructions on how to proceed correctly during calibration.

 

Perform calibration with the UWS WaterBoy app

  • To do this, switch to the desired tab (μS or pH)
  • If the calibration is invalid or has expired, the message ‘Device must be calibrated’ is highlighted in red at the bottom.
  • If the calibration has expired, please recalibrate the respective measuring probe.
  • Click on ‘Calibrate’ for step-by-step instructions on calibration.
  • Important: As soon as the device has been calibrated to a calibration solution, the app jumps out of calibration again. To calibrate another point, click on ‘Calibrate’ again and follow the instructions.
  • The date of the calibration and the calibrated points are displayed after successful calibration.

A detailed description of how to calibrate the UWS WaterBoy without the app can be found here in the instructions.

Calibration of pH and conductivity
play
Play video

The following calibration liquids are available:

Calibration fluid Conductivity
Article Article content Species number
Conductivity solution 84µS / cm 120 ml bottle 100030-8
Conductivity solution 1413 µS/cm 25 bags of 20 ml 100030-2
Conductivity solution 1413 µS/cm 500 ml bottle 100030-25
Calibration liquid pH
Buffer solution pH 4.01 25 bags of 20 ml 100030-3
Buffer solution pH 4.01 500 ml bottle 100030-30
Buffer solution PH 7 25 bags of 20 ml 100030-1
Buffer solution PH 7 500 ml bottle 100030-15
Buffer solution pH 10.01 25 bags of 20 ml 100030-7
Buffer solution pH 10.01 500 ml bottle 100030-70
Cleaning and storage solution
Storage solution for electrodes 25 ml bottle 100125
Storage solution for electrodes 230 ml bottle 100145
Storage solution for electrodes 500 ml bottle 100135
Cleaning solution for electrodes 25 bags of 20 ml 100030-6
Cleaning solution for electrodes 500 ml bottle 100136
Replacement set “WaterBoy” measuring case 10 pcs. 100153-1

We also offer the following measuring and test equipment:

article Item number
Measuring set for measuring water hardness 100032
Chloride test kit 100130
Test cutlery iron 100131
Nitrate test kit 100133
Sulphate test kit 100134
Measuring cup 50 ml 100030-4

Contact
find on site:

Capacity calculator for reprocessing units

With our capacity calculator you can easily and quickly calculate the filling capacity of your unit as well as the expected required amount of Vadion pH-Control mixed bed resin. Please select whether it is a new or existing unit.
UWS Füllgerät
Mischbetttauscher
°dH des Wassers
Heizwasser in Liter
Füllkapazität Ihres Gerätes:
Füllwasser Kapazität ca.
20.700 l
Benötigte Gesamtmenge Vadion in Liter ca.
0,11 l
Contact us
On June 20 we have an extra day off to make it a long weekend